Product |
Seplife® LXMS 30QN |
Appearance |
White to light yellow spherical beads |
Type |
Strong anion exchange resin |
Matrix |
Polystyrene/DVB |
Ligand |
Quaternary Amine |
Ion exchange capacity (mmol/ml) |
0.07-0.10 |
Particle size (μm) |
30±3 |
Pore size (nm) |
55±10 |
pH stability |
2-12 (operation), 1-14 (CIP) |
Chemical stability |
All common ion exchange buffers. |
Pressure flow rate (cm/h)* |
100-900 |
Dynamic capacity (mg/ml, BSA)** |
≥40 |
Maximum Pressure resistance |
3.0 MPa / 30 Bar |
Shipped as |
20% ethanol slurry |
The Seplife ® LXMS 30QN resin is supplied as a 50-60% (v/v) slurry in 20% ethanol solution. The use of a high ionic strength mobile phase for resin packing into the chromatographic column (including cleaning and disinfection) is preferred ; it is recommended to use 0.5 M NaCl; 0-20% ethanol solution can also be suitable. The recommended slurry concentration for column packing is 50-60% (v/v).
Equilibrate the chromatographic column with mobile phase of 0.5M NaCl solution at a flow rate of 100-120cm/h for 5-10CV. Test the column efficiency using a mobile phase of 0.5M NaCl solution and injecting 0.5-1% column volume of 2M NaCl solution at 100-120cm/h flow rate. Using the conductivity detector, record the chromatogram and calculate the peak asymmetry and the theoretical plate number. Typically, the number of plates is ≥8000/m, and the asymmetry factor is 0.8-1.5.
The packed columns should be rinsed with a minimum 5 CV of buffer.
After packing the column, equilibrate with the mobile phase first, with 5-10 column volumes, and control the flow rate at 120-300cm/h until the conductivity and pH of the flow-through remain unchanged before feeding the sample.
If the sample is in solid form, it can be prepared by dissolving in the equilibration buffer; a low-concentration sample solution should be concentrated in advance as much as possible; a high concentration sample solutions should be diluted with equilibration buffer. To avoid clogging of the column, samples should be processed by centrifugation or membrane filtration. The feed amount is calculated according to the capacity of the resin and the content of the target molecule in the feed solution. Before loading, make sure that the sample buffer should be as consistent as possible with the equilibration buffer.
After loading the sample, continue rinsing with equilibration buffer until the baseline is stable. According to the actual situation, elute the samples adsorbed on the resin sequentially by increasing the salt concentration or changing the pH of the mobile phase.
Regular Cleaning-In-Place (CIP) can prevent column fouling, and help to maintain the capacity and resolution of the chromatographic media. Specific CIP methods and the frequency of CIP need to be designed for each process according to the type of contamination. The The recommended regeneration and CIP method is as follows: rinse the column up-flow with 5 CV of 1-2 M NaCl followed by 5 CV of 0.5-1 M NaOH.
Chromatography resins in bulk that are not for immediate use should be stored in 20% ethanol at 2-30 °C. The column packed with Seplife ® LXMS 30QN , after regeneration, CIP and sanitization should be stored in a buffer solution containing 20% ethanol preferable at neutral pH.
Product Name |
References |
Pack Size |
Seplife® LXMS 30QN |
PS30333X(30)1-1 |
25ml |
PS30333X(30)1-2 |
100ml |
|
PS30333X(30)1-3 |
500ml |
|
PS30333X(30)1-4 |
1L |
|
PS30333X(30)1-5 |
5L |
|
PS30333X(30)1-6 |
10L |
Production date: See label
Expiry date: 5 years, under proper storage conditions