Product |
Seplife® RP LX-21SS (L) |
Appearance |
White spherical beads |
Matrix |
Poly divinylbenzene |
Particle size (μm) |
100-300 |
Uniformity Coefficient |
≤1.3 |
pH Limits, Stability |
1-14 |
Pore Volume |
1.70±0.40 |
Specific Surface Area (m2/g) |
≥600 |
Pore Radius (Å) |
150-500 |
Shipped as |
20% Ethanol Slurry |
(6) Lock the column plunger in the compressed position so that the column can be operated in the Static Axial Compression (SAC) mode.
(7) The packed column is now ready for use. It can be used while still assembled on the packing station or it can be undocked for use in a purification facility.
After packing, clean the chromatographic column with 3-5 CV of 100% methanol or 80% acetonitrile solution. The flow rate should be controlled at 120-180 cm/h to balance and perform column efficiency test.
The test method for column efficiency of RP chromatography columns is as follows:
Mobile phase: 100% methanol or 80% acetonitrile solution
Linear flow rate: 120-180 cm/h
Sample: 1:9 (v:v) Acetone in 100% methanol or 80% acetonitrile
Loading volume: 1% of column volume;
Detection: UV @ 254 or 280 nm;
The prep-HPLC system geometry, including dead volume, will significantly affect the plate count determination.
Equilibrate with the mobile phase for 3-4 CV, and control the flow rate at 120-180 cm/h until the conductance and pH of the flow-through remain unchanged before loading the sample.
The solid sample can be prepared by dissolving in the equilibrium solution. Low-concentration sample solutions can be concentrated in advance as much as possible while too high concentration sample solution can be diluted with the equilibration solution. To avoid clogging of the column, samples should be processed by centrifugation or membrane filtration. The feed amount is calculated according to the capacity of the resin and the content of the target protein in the feed solution. Before loading, ensure the sample buffer is as consistent as possible with the equilibration solution.
Use 2-10 CV of methanol, ethanol, acetonitrile, acetone, etc. (aqueous) solution to elute; use acid, caustic or buffer to adjust the pH or use a combination both to elute the molecules of interest.
First use acetonitrile, methanol, ethanol, acetone, NaOH in ethanol or other solvents to wash the column (3-4 CV) according to the operating flow rate, and then use the equilibration solution to rinse (3-4 CV).
Product Name |
References |
Pack Size |
Seplife® LX-21SS(L) |
PS00052L2-1 |
25ml |
PS00052L2-2 |
100ml |
|
PS00052L2-3 |
500ml |
|
PS00052L2-4 |
1L |
|
PS00052L2-5 |
5L |
|
PS00052L2-6 |
10L |
Production date: See label
Expiry date: 5 years, under proper storage conditions